Rotating-arm method for low speed calibration of hot-wire probes in water applications

Yükleniyor...
Küçük Resim

Tarih

Yazarlar

Özkan, Musa

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Elsevier

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

An instrument for calibration of hot-wire probes at very low velocities in water applications is introduced.The technique is based on moving a hot-wire probe by means of a rotating-arm device in stillwater. This simple in situ calibration method is principally beneficial for the investigations of threedimensionalboundary layers which are encountered for instance over rotating disks. In addition to thisproposed technique, an existing alternative method in which the calibration is performed against aknown reference velocity over a flat surface of a rotating disk is conducted. The comparison of both methodsshow that the rotating-arm technique is superior to that existing alternative flat-surface calibrationmethod.

Açıklama

Özkan, Musa (Bilecik, Author)

Anahtar Kelimeler

Rotating Arm, Hot-wire Probes, Low Speed Calibration, Three-dimensional Boundary Layers, Rotating-disk Flow

Kaynak

Measurement

WoS Q Değeri

Scopus Q Değeri

Cilt

145

Sayı

Künye

Özkan, M. (2019). Rotating-arm method for low speed calibration of hot-wire probes in water applications. Measurement.