Magnetometric Sensitivity and Spin Relaxation of Buffered Gas Cell for Cardiomagnetometer
Yükleniyor...
Tarih
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Ieee
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
We investigated a diode-pumped Cs-based Mx type atomic magnetometer developed aiming the detection of low magnetic fields of the human heartbeat. In this study, we characterized and optimized the rf coil voltage and the laser power to achieve the minimum magnetic resonance linewidth for the best sensitivity. The measured half-width of an 18 Hz at half maximum and a signal-to-noise ratio of 22800 with shot-noise-limited were obtained. The sensitivity of the magnetometer was shown to be around 228 fT/vHz which is under the required limit for the cardiomagnetometer applications.
Açıklama
IEEE International Symposium on Medical Measurements and Applications (IEEE MeMeA) -- JUN 26-28, 2019 -- Istanbul, TURKEY
Anahtar Kelimeler
high precision magnetometry, optically detected magnetic resonance, cardio magnetometer
Kaynak
2019 Ieee International Symposium on Medical Measurements and Applications (Memea)